beautypg.com

Cilxs_configurepulse – AMETEK Ls Series Driver Manual User Manual

Page 70

background image

IVI Driver Manual

Lx / Ls Series

66

ciLxs_ConfigurePulse


ViStatus ciLxs_ConfigurePulse (ViSession instrumentHandle,
ViInt32 count, ViReal64 width,
ViReal64 period);

Purpose

This function configures the generation of output pulses.

Parameter List

instrumentHandle

Variable Type ViSession

The ViSession handle that you obtain from the ciLxs_init or
ciLxs_InitWithOptions function. The handle identifies a particular
instrument session.

Default Value: None


count

Variable Type ViInt32

Pass the number of pulses that are output when a triggered output
transient occurs. The driver sets the CILXS_ATTR_TRIGGER_PULSE_COUNT
attribute to this value.


Valid Range: 1 to 2147483647

Default Value: 1

width

Variable Type ViReal64

Pass the the width of a transient output pulse. The driver sets the
CILXS_ATTR_TRIGGER_PULSE_WIDTH attribute to this value.

Units: Seconds

Valid Range:

3-phase models: 0 to 1.07533e6
1-phase models: 0 to 4.30133e5

Default Value: 0.01667

period

Variable Type ViReal64

Pass the period of a triggered output transient. The driver sets the

This manual is related to the following products: