beautypg.com

Cilxs_configuretriggerfunctionlist – AMETEK Ls Series Driver Manual User Manual

Page 108

background image

IVI Driver Manual

Lx / Ls Series

104

ciLxs_ConfigureTriggerFunctionList


ViStatus ciLxs_ConfigureTriggerFunctionList (ViSession instrumentHandle,
ViChar _VI_FAR
functionsList[]);

Purpose

This function configures the sequence of the waveform shape entries.

Parameter List

instrumentHandle

Variable Type ViSession

The ViSession handle that you obtain from the ciLxs_init or
ciLxs_InitWithOptions function. The handle identifies a particular
instrument session.

Default Value: None


functionsList

Variable Type ViChar[]

The sequence of function shapes list points.

The following values may be specified:
"SINUSOID", "SQUARE", "CSINUSOID" and user define function names.

The function names are separated by comma.

Return Value

Returns the status code of this operation. The status code either
indicates success or describes an error or warning condition. You
examine the status code from each call to an instrument driver
function to determine if an error occurred.

To obtain a text description of the status code, call the
ciLxs_error_message function. To obtain additional information about
the error condition, call the ciLxs_GetError function. To clear the
error information from the driver, call the ciLxs_ClearError
function.

The general meaning of the status code is as follows:

Value Meaning
-------------------------------
0 Success
Positive Values Warnings
Negative Values Errors

This driver defines the following status codes:

This manual is related to the following products: