Cilxs_configuretriggerdelay – AMETEK Ls Series Driver Manual User Manual
Page 82

IVI Driver Manual
Lx / Ls Series
78
ciLxs_ConfigureTriggerDelay
ViStatus ciLxs_ConfigureTriggerDelay (ViSession instrumentHandle,
ViReal64 triggerDelay);
Purpose
This function configures the time delay between the detection of a
trigger signal and the start of any corresponding trigger action.
Parameter List
instrumentHandle
Variable Type ViSession
The ViSession handle that you obtain from the ciLxs_init or
ciLxs_InitWithOptions function. The handle identifies a particular
instrument session.
Default Value: None
triggerDelay
Variable Type ViReal64
Pass the time delay between the detection of a trigger signal and the
start of any corresponding trigger action. After the time delay has
elapsed, the trigger is implemented unless the trigger system is also
waiting for a sync signal that has been
specified by ciLxs_ConfigureSynchronizationPhase. The driver sets
the CILXS_ATTR_TRIGGER_DELAY attribute to this value.
Units: seconds
Defined Values:
3-phase models: 0 to 1.07533e6
1-phase models: 0 to 4.30133e5
Default Value: 0.0 second
Return Value
Returns the status code of this operation. The status code either
indicates success or describes an error or warning condition. You
examine the status code from each call to an instrument driver
function to determine if an error occurred.
To obtain a text description of the status code, call the
ciLxs_error_message function. To obtain additional information about
the error condition, call the ciLxs_GetError function. To clear the
error information from the driver, call the ciLxs_ClearError