4.8 Setting the Power-on Initialization Values

The power source is shipped with default factory settings when the unit is powered up. The factory
settings are:

Parameter

Factory default setting

Voltage 0.0

Volt

Voltage Range

150 Volt Range

Frequency 60

Hz

Current limit

Maximum available current for selected V Range.

Output state

OFF

Local / Remote State

Local. Front panel unlocked.

Table 4-3: Factory Default Power on Settings

It is possible to change the power on initialization values in one of two ways:

1. Using the IEEE-488, LAN or USB bus interface.

2. Using the front panel.

To change the power on initialization values from the front panel, proceed as follows:

1. Set the AC power source output parameters from the front panel as you want to power up the

unit.

2. Save this setting to setup register 0 from the REGISTERS menu.

3. Select the CONFIGURATION menu and move to the POWER ON field.

4. Change the POWER ON field to REG0.

5. This will recall the settings contained in register 0 at power up.