AMETEK Compact IX 2253 User Manual

Page 153

background image

User Manual

2253i / 2253iX

AMETEK California

Instruments

153

Number

Message String

Cause

Remedy

lost"

contents lost

-315 "Configuration

memory

lost"

Hardware configuration
settings lost.

Contact CI service department at

[email protected]

to

obtain instructions on restoring
configuration data.

-330

"Self-test failed"

Internal error

Contact CI service department at

[email protected]

-350

"Queue overflow"

Message queue full.

Too many messages. Read status using
SYST:ERR query until 0, "No Error" is
received indicating queue empty.

-400

"Query error"

Unable to complete query. Check programming manual for correct

query format and parameters

-410

"Query INTERRUPTED"

Query issued but
response not read.

Check application program for correct
flow. Response must be read after each
query to avoid this error.

-420 "Query

UNTERMINATED"

Query incomplete.

Check for terminator after query
command.

-430 "Query

DEADLOCKED"

Query cannot be
completed

Check application program for multiple
queries

-440 "Query

UNTERMINATED"

Query incomplete.

Check for terminator after query
command.

0

"No error"

No errors in queue

40

"Voltage self test error,
output 1”

No. 1/A amplifier in Master
source has no output
during Self-test.

Contact CI service department at

[email protected]

41

"Voltage self test error,
output 2”

No. 2/B amplifier in Master
source has no output
during Self-test. Three
phase models only.

Contact CI service department at

[email protected]

42

"Voltage self test error,
output 3”

No. 3/C amplifier in
Master source has no
output during Self-test
Three phase models only.

Contact CI service department at

[email protected]

43

"Current self test error,
output 1”

No. 1/A amplifier in Aux.
Source has no output
during Self-test.

Contact CI service department at

[email protected]

44

"Current self test error,
output 2”

No. 2/B amplifier in Aux.
Source has no output
during Self-test. Three
phase models only.

Contact CI service department at

[email protected]

45

"Current self test error,
output 3”

No. 3/C amplifier in Aux.
Source has no output
during Self-test. Three
phase models only.

Contact CI service department at

[email protected]

403

"CAL not enabled"

No password entered for
calibration

Enter correct CAL password.

801

"Output volt fault"

- Output voltage does not
match program value
when ALC is on.
- Over load
- Voltage kick-back
- No output voltage

Load exceeds current limit and unit is in
Constant Voltage (CV) mode of
operation.
- Reduce load or increase CL setting.
Output voltage is driven above
programmed voltage by external