7.4 Self test
An internal self-test can be performed over the bus by sending the *TST? query command. The self-
test will run until the first error is encountered and terminate. The response to the query will either be
the first error encountered or 0 if no error was found (Self-test passed).
To execute a self-test, the IEEE-488, LAN, RS232 or USB interface must be used. The iXCGui
command self-test window can be used to send the *TST? Command. See the section 9 for possible
self test error codes and messages.
Note that the internal selftest is more comprehensive than the power on memory test that always runs
when the unit is turned on. The memory test only checks the integrity of the controller and its non-
volatile memory content. The self test runs an actual power test on all amplifiers.
7.5 Advanced Troubleshooting.
CAUTION:
VOLTAGES UP TO 230 VAC AND 500 VDC ARE PRESENT IN CERTAIN
SECTIONS OF THIS POWER SOURCE.
WARNING:
THIS EQUIPMENT GENERATES POTENTIALLY LETHAL VOLTAGES.
DEATH ON CONTACT MAY RESULT IF PERSONNEL FAIL TO OBSERVE SAFETY
PRECAUTIONS. DO NOT TOUCH ELECTRONIC CIRCUITS WHEN POWER IS APPLIED
Switch Off Units
Switch off each unit at the circuit breaker on the front panel as well as removing the input power from
the unit.
WARNING:
Wait 10 minutes for all internal capacitors to discharge.
Removing Cover
Remove the screws securing the top cover and remove it.
Initial Inspection
Make a visual inspection of the unit and ensure all the connectors are properly mated and there are no
loose wires.