0V for 2 seconds.
100V for 8 seconds.
The
key (backspace) will terminate the test at any time.
OVER FREQUENCY
This test will change the output frequency in the sequence shown in Table 10-4 before returning to the
steady state frequency.
The
key (backspace) will terminate the test at any time.
Revision
D
E
F
F 60Hz only
FREQ
Time
FREQ
Time
FREQ
TIME
FREQ
TIME
Seq1
480Hz 5sec. 480Hz 5sec. 480Hz 5sec 61Hz 7sec
Seq2
420Hz 5sec 420Hz 9sec 420Hz 5sec 60.5Hz 8sec
Table 10-4: Abnormal Over Frequency
After this sequence, a 5 second delay will be inserted at the nominal test frequency. The
key
(backspace) will terminate the test at any time.
UNDER FREQUENCY
This test will change the output frequency in the sequence shown in Table 10-5 before returning to
steady state frequency.
The
key (backspace) will terminate the test at any time.
Revision
D
E
F
F 60Hz only
FREQ
Time
FREQ
Time
FREQ
TIME
FREQ
TIME
Seq1
0 5sec. 0Hz 7sec. 0Hz 7sec 0Hz 7sec
Seq2
375Hz 5sec 380Hz 7sec 380Hz 3sec 59.5Hz 8sec
Table 10-5: Abnormal Under Frequency
After this sequence, a 5 second delay will be inserted at the nominal test frequency. The
key
(backspace) will terminate the test at any time.