 0V for 2 seconds.
 100V for 8 seconds.

The

 key (backspace) will terminate the test at any time.

OVER FREQUENCY

This test will change the output frequency in the sequence shown in Table 10-4 before returning to the
steady state frequency.

The

 key (backspace) will terminate the test at any time.

Revision

D

E

F

F 60Hz only

FREQ

Time

FREQ

Time

FREQ

TIME

FREQ

TIME

Seq1

480Hz 5sec. 480Hz 5sec. 480Hz 5sec 61Hz 7sec

Seq2

420Hz 5sec 420Hz 9sec 420Hz 5sec 60.5Hz 8sec

Table 10-4: Abnormal Over Frequency

After this sequence, a 5 second delay will be inserted at the nominal test frequency. The

 key

(backspace) will terminate the test at any time.

UNDER FREQUENCY

This test will change the output frequency in the sequence shown in Table 10-5 before returning to
steady state frequency.

The

 key (backspace) will terminate the test at any time.

Revision

D

E

F

F 60Hz only

FREQ

Time

FREQ

Time

FREQ

TIME

FREQ

TIME

Seq1

0 5sec. 0Hz 7sec. 0Hz 7sec 0Hz 7sec

Seq2

375Hz 5sec 380Hz 7sec 380Hz 3sec 59.5Hz 8sec

Table 10-5: Abnormal Under Frequency

After this sequence, a 5 second delay will be inserted at the nominal test frequency. The

 key

(backspace) will terminate the test at any time.