beautypg.com

AMETEK Compact IX 2253 User Manual

Page 155

background image

User Manual

2253i / 2253iX

AMETEK California

Instruments

155

Number

Message String

Cause

Remedy

826

“Peak current prot trip”

Peak current limit
exceeded.

Peak current exceeded. Could be caused
by switching EUT on or off.

827

“Frequency error”

Frequency error during
self-test.

Correct frequency was not measured
during self-test. May be result of 801
error.

828

“Phase error”

Self test error phase angle Correct phase angle was not measured

during self-test. May be result of 801
error.

829

"Dc component exceed
limit"

DC value exceed limit for
the voltage mode setting.

Select AC+DC mode instead of AC
mode.

830

"Lan connection missing” LAN Cable is not

connected

Make sure the LAN cable is connected to
a network.

831

"Duplicate IP,Set to
DHCP”

It detect a duplicate IP on
the network.

Change IP or Cycle power to connect to
DHCP or Auto IP.

832

"Duplicate IP,REM set
reject”

The IP set from remote
interface is rejected.

The IP is rejected use different IP. It is a
duplicate IP or out of reach IP.

833

“Duplicate IP,LOC set
rejec”

The IP set from front
panel keypad is rejected

The IP is rejected use different IP. It is a
duplicate IP or out of reach IP.

834

"Out of reach IP"

The IP set is out of reach

Use an IP within the sub net.

Table 16-1: Error Messages