AMETEK Compact IX 2253 User Manual

Page 74

background image

User Manual

2253i / 2253iX

AMETEK California

Instruments

74

4.2.10 CALIBRATION Menus

The measurement calibration menu can be used to perform routine calibration of the internal
measurement system. The recommended calibration interval is 12 months. To enter the calibration
screens, the calibration password must be entered first.

Note: Refer to chapter 6 for details on routine calibration procedures and equipment requirements. Do
not attempt calibration without consulting the user manual.

This menu also contains the LCD viewing angle adjustment.

ENTRY

DESCRIPTION

VIEW ANGLE

-10 to +10

LCD viewing angle adjustment.

CAL PWORD

V range

Calibration password required to access all
calibration screens. The calibration password
is the high voltage range value. [300 ] The
password can be entered using the keypad or
shuttle followed by the ENTER key.

Measurement Calibration Screens

MVOLT FS

± 9999

Calibration coefficient for full-scale voltage
measurement.

MCURR FS

± 9999

Calibration coefficient for full-scale current
measurement.

Output Calibration Screens

VOLT FS

± 9999

Calibration coefficient for voltage output.

PHASE OFST

0.0 - 360.0

Phase offset calibration for units equipped
with the –LKS option. Determines phase offset
from –LKM master clock.