VOLT FREQ MIN

Standard/Group

RTCA

A(CF)

A(NF)

A(WF)

Voltage

1

100 100 100 100

3

101.5 101.5 101.5 101.5

Frequency

380 390 360 360

Standard/Group

Group1

Group2

Group3

Voltage 1 104

104

104

3

105.5 105.5 105.5

Frequency 390

360

360

Table 9-1: Normal Voltage and Frequency minimum

Standard/Group

RTCA

A(CF)

A(NF)

A(WF)

Voltage

1

122 122 122 122

3

120.5 120.5 120.5 120.5

Frequency

420 410 650 800

Standard/Group

Group1

Group2

Group3

Voltage 1 122

122

122

3

120.5 120.5 120.5

Frequency 410

650

800

Table 9-2: Normal Voltage and Frequency Maximum

This test will set the voltage and frequency to levels defined by Table 9-1. The test will last for 30 minutes.
The test will be repeated, except group1, using the Voltage setting from Table 9-2 and the frequency from
Table 9-1. The

 key (backspace) will terminate the test at any time.

VOLT FREQ MAX

This test will set the voltage and frequency to levels defined by Table 9-1. The test will last for 30 minutes.
The test will be repeated, except group1, using the Voltage setting from Table 9-1 and the frequency from
Table 9-2. The unselected phases will remain at 115 volts. The BACK key (backspace) will terminate
the test at any time.