AMETEK Compact IX 2253 User Manual
Page 116

User Manual
2253i / 2253iX
AMETEK California
Instruments
116
VOLT FREQ MIN
Standard/Group
RTCA
A(CF)
A(NF)
A(WF)
Voltage
1
100 100 100 100
3
101.5 101.5 101.5 101.5
Frequency
380 390 360 360
Standard/Group
Group1
Group2
Group3
Voltage 1 104
104
104
3
105.5 105.5 105.5
Frequency 390
360
360
Table 9-1: Normal Voltage and Frequency minimum
Standard/Group
RTCA
A(CF)
A(NF)
A(WF)
Voltage
1
122 122 122 122
3
120.5 120.5 120.5 120.5
Frequency
420 410 650 800
Standard/Group
Group1
Group2
Group3
Voltage 1 122
122
122
3
120.5 120.5 120.5
Frequency 410
650
800
Table 9-2: Normal Voltage and Frequency Maximum
This test will set the voltage and frequency to levels defined by Table 9-1. The test will last for 30 minutes.
The test will be repeated, except group1, using the Voltage setting from Table 9-2 and the frequency from
Table 9-1. The
key (backspace) will terminate the test at any time.
VOLT FREQ MAX
This test will set the voltage and frequency to levels defined by Table 9-1. The test will last for 30 minutes.
The test will be repeated, except group1, using the Voltage setting from Table 9-1 and the frequency from
Table 9-2. The unselected phases will remain at 115 volts. The BACK key (backspace) will terminate
the test at any time.