VOLT FREQ MAX

This test will set the voltage and frequency to a level defined by Table 9-9. The test will last for 30
minutes. The test will be repeated using the voltage from Table 9-8 and frequency from Table 9-9.
The BACK key (backspace) will terminate the test at any time.

VOLT UNBALANCE

Standard/Group

RTCA

A(CF)

A(NF)

A(WF)

Voltage offset

8

8

8

10

Frequency 400 360/440

360/650

360/800

Standard/Group

Group1

Group2

Group3

Voltage offset

8

8

12

Frequency 400 360/650

360/800

Table 9-10: Emergency Voltage Unbalance

This test will change the output voltage for phase A and B to 122V and phase C to a voltage lower by
a value specified by an offset. Refer to Table 9-10 for the offset value and the Frequency. The test
will repeat with the same frequency and phase A and B volt is set to 100V and phase C set to a higher
voltage specified by the offset. The test will last 30 minutes. The test will be repeated for a second
Frequency if applicable. The test can be terminated at any time.

The BACK key (backspace) will terminate the test at any time.