138

 420Hz for 4 sec.
 410Hz for 5 sec.
 407Hz for 4 sec.

For 60Hz only:

 61Hz for 0.5 sec.
 60.5Hz for 0.5 sec.

After this sequence, a 5 second delay will be inserted at the nominal test frequency. The

 key

(backspace) will terminate the test at any time.

LOW FREQUENCY TRANSIENT

This test will change the output frequency in the following sequence:

For 400Hz and VFREQ:

 375Hz for 1 sec.
 380Hz for 4 sec.
 390Hz for 5 sec.
 393Hz for 4 sec.

For 60Hz only:

 59Hz for 0.5 sec.
 59.5Hz for 0.5 sec.

After this sequence, a 5 second delay will be inserted at the nominal test frequency. The

 key

(backspace) will terminate the test at any time.