FREQUENCY

This test will change the output frequency in the sequence shown in Table 10-2.

SEQUENCE

FREQUENCY

TIME

400Hz

VFREQ

60 Hz

1 393

360

59

1

minute

2 407

800

61

1

minute

3 400

SSF

60

1

minute

Table 10-2: Steady state frequency

The

 key (backspace) will terminate the test at any time.

VOLT UNBALANCE

This test will change the output voltage for the selected phase only in the following sequence:

 112V for 1 minute.
 118V for 1 minute.
 115V for 1 minute.

The test will be repeated on three phase systems to include all three phases if the coupling is set to all.

The

 key (backspace) will terminate the test at any time.

PHASE DIFFERENCE

This test applies to three phase systems only. The phase angle for the selected phase will change
relative to phase A in the following sequence:

If phase B is selected:

 236 for 1 minute.
 244 for 1 minute.
 240 for 1 minute.

If phase C is selected:

 116 for 1 minute.
 124 for 1 minute.
 120 for 1 minute

VOLTAGE MODULATION

This test will vary the output voltage by

2.5V rms over a period of one second. The test will last for 2

minutes. The

 key (backspace) will terminate the test at any time.

FREQUENCY MODULATION

REVISION

D

E

F (400Hz /VFREQ)

F (60HZ)

MODULATION

7Hz

4Hz

4Hz

0.5Hz

Table 10-3: Frequency Modulation

This test will vary the output frequency as defined by Table 10-3 over a period of one minute. The test
will last for 4 minutes. The

 key (backspace) will terminate the test at any time.