AMETEK Compact IX 2253 User Manual
Page 88

User Manual
2253i / 2253iX
AMETEK California
Instruments
88
Figure 4-12: Sample Transient Output Sequence
This output can be accomplished using the following transient list.
Step #
(data point)
Volt VSlew
Frequency
FSlew Dwell
0 70.00 MAX 360.0 MAX 0.100
1 110.00 100.0 440.0 MAX 0.900
2 130.00 MAX 240.0 800.00 0.250
3 90.00 53.3 240.0 MAX 0.750
4 90.00 MAX 460.0 MAX 1.000
5 110.00 MAX 400.0 MAX 0.800
6 88.00 MAX 400.0 MAX 0.200
7 110.00 MAX 400.0 MAX 1.000
Table 4-2: Sample Transient List
4.7.7 Waveforms Function List
The FUNCTION field available in each transient list event setup menu may be used to dynamically
switch waveforms during transient execution. This allows different waveforms to be used during
transient execution. Waveforms may be switched without the output of the source being turned off. For
three phase configurations, each phase has its own waveform list so different waveforms may be
programmed on different phases during transient execution.
Figure 4-13 illustrates the concept of using different waveforms at different steps in a transient list. In
this case, the change was programmed to occur at the zero crossing. Any phase angle can be used to
start the transient execution however. To keep the phase angle synchronization, the dwell times have