9.5 AC Test Mode

Following sections cover testing in AC output mode.

9.5.1 Normal State tests

Scroll to the NORMAL STATE entry using the up and down cursor keys. Press the ENTER key to
select the NORMAL STATE screens. The screen will appear as shown in Figure 9-3.

Figure 9-3: Normal state screens

The DO160 NORMAL screens have the following tests:

1

VOLT FREQ MIN

2

VOLT FREQ MAX

3 VOLT

UNBALANCE

4

WAVEFORM DISTORTION

5 VOLT

MODULATION

6 FREQ

MODULATION

7 POWER

INTERRUPT

8 VOLTAGE

SURGE

9

FREQ TRANSIENT (group 1/A(CF))

FREQ VARIATION (group 2 & 3/A(NF) & A(WF))

The above tests can be selected by scrolling to the highlighted selection using the up and down key
and the ENTER key to start the selected test. For some of these tests, numeric data entry may be
required to define the test number or the modulation rate.