beautypg.com

AMETEK Compact IX 2253 User Manual

Page 70

background image

User Manual

2253i / 2253iX

AMETEK California

Instruments

70

4.2.8 REGISTERS Menu

The registers menu provides access to the non-volatile setup storage of the power source. A total of
16 front panel setups can be stored in registers numbered from 0 through 15. Each register except
register 0 can hold the complete front panel setup, including the programmed transient list. This allows
for quick recall of different setups and transient programs.

Register 0 is reserved to be used as the power-on setting as assigned by the user. To have the power
source start in a specific setting, save the desired setting to Register 0 and assign register zero as the
power-on default in the CONFIGURATION menu. Alternatively, the power source can be set to power
up with the RST factory default settings. See 4.9 for factory default settings.

ENTRY

DESCRIPTION

SAVE

REG 0 – 15

Saves the selected setup and transient list
from memory. (Setup only for Reg 0) The
shuttle knob may be used to scroll through the
available list of setup register numbers.
Use the ENTER key to perform the save
operation.
Register 0 can be assigned as the power-on
state setup from the CONFIGURATION menu.
A valid setup must be saved in REG0 to do
so.

Note

that REG0 only saves the setup, not the

transient list. All other registers also save the
transient list.

RECALL

REG 0 – 15

Recalls the selected setup and transient list to
memory. (Setup only for Reg 0) The shuttle
knob may be used to scroll through the
available list of setup register numbers.
Use the ENTER key to perform the recall
operation.
Register 0 can be assigned as the power-on
state setup from the CONFIGURATION menu.
A valid setup must be saved in REG0 to do
so.

Note

that REG0 only saves the setup, not the

transient list. All other registers also save the
transient list.