4.2.9 CONFIGURATION Menu
The configuration menu may be used to configure various aspects of the instrument such as the serial
port, IEEE-488/GPIB address, LAN settings (units with -LAN option only) and the power-on settings of
the supply.
ENTRY
DESCRIPTION
ADDRESS
0 - 31
Sets the selected IEEE / GPIB bus address
for the optional IEEE/GPIB interface.
Factory default is address 1. The shuttle knob
or the keypad can be used to set a value from
0 through 31. Do not use address 0 as this
address is typically reserved for the GPIB
controller.
BAUD RATE
9600
19200
38400
57600
115200
230400
460800
Sets the baud rate for the USB, RS232 and/or
LAN (Ethernet) communications port. Factory
default is 460800 baud. Available settings are
9600 through 460800 baud.
Note:
For USB and LAN use, you must set
the baud rate to 460800. The shuttle knob can
be used to scroll through these selections.
PON STATE
REG0
RST
Determined power on state. This setting
selects either non-volatile REG0 to be recalled
automatically at power-on or factory default
(RST). Factory default is RST, which recalls
the factory settings.
Note that to use REG0 for power-on default,
the contents of the register must be
programmed first. See section 4.2.8. If an
empty register is selected, the power source
will revert back to RST (factory setting).
CONTROL MAST
AUX
This is an information-only field that displays
the controller operation mode. For a single
stand-alone iX unit, the mode is always MAST
(Master).
LANGUAGE SCPI
ABLE
Displays the active programming command
language syntax selection. Default for
compact i/iX series is SCPI (Standard
commands for Programmable Instruments). If
the –ABL option is installed, the ABLE (Atlas
Based Language Extension) syntax can be