Kipp&Zonen BSRN Scientific Solar Monitoring System User Manual
Page 123
111
Cavity size; CW L tolerance
3-cavity, Ø25 m m ; ±2 nm
Side-band blocking
OD4, UV to 1200 nm
Detector type; active area
UV si-photodiode; 33 m m 2
Sensitivity gain setting x 4 channels
high/low by jum per; trim via m ulti-turn pot
Output signal x 4 channels
-0.05 to 4.5 VDC m ax.
Resolution
<0.005OD (Langley m ethod)
Response tim e
0.2 s to 99%
Operating tem perature
-30 °C to +70 °C
Power supply requirem ent
5.5 to 14.5 VDC, 20 m A
Tem perature output
10 m V °C (0.20 V = 20 °C)
-1
Desiccant
silica gel (visible through front window)
Interface lead
8-core shielded cable, 5 m (free end is unterm inated)
Mounting m ethod
Ø25 m m (1") fem ale clam p, with adjustable alignm ent
W eight
1.25 kg (excluding 0.5 kg lead)
B 5.7
Carter-Scott Design Middleton SP01 Sunphotom eter
The SP01 version has fixed field stops and operates in sun photom eter m ode only.
•
Four narrow-bandwidths m easured sim ultaneously
•
Detectors and filters held at constant tem perature to enhance stability
•
Low- noise signal am plifiers with negligible zero offset drift
•
Three gain selections accom m odate a wide range of signal strengths
•
Motorised field stop on SP01-A can pause for signal zero check
•
W eather-proof alum inum housing
Specifications
(A) Spectro-Pyrheliom eter Tube
Dim ension
48 x 312 nm
Field of view:
sun photom eter field stop
2.4° (1.2° opening)
solar aureole field stop
opening angle 3° - 5°, centre-blocked
(SP01-A only)
Slope angle (with 2.4° field stop)
0.9° (lim it angle = 1.5°)
Filter wavelength and bandpass
368BP5nm ; 412BP5; 500BP5; 862BP5
(Optional filters)
675BP5; 778BP5; 812BP5
Filter size and CW L tolerance
Ø25 m m , ± 2 nm
Sensitivity gain selection (by jum per)
low (= m edium x 0.2); m edium ; high (= m edium x 5)
Output signal
-0.05 to +8.0 VDC m ax
Response tim e
2 s to 99%