CALibration:WAVelength

Description This command introduces an offset into the FLS-2600B to

correct a wavelength deviation as measured by a wavelength

meter.

Syntax

CAL:WAV

Parameters The parameter is a value indicating the

wavelength in nm.

Example

CAL:WAV 1560.000

Note

This function cannot be performed if the application is

performing a sweep.
The difference between the wavelength and the measured

wavelength cannot be greater then ± 0.200 nm or ± 0.0250 THz.
EXFO recommends using 1560 nm as the wavelength