beautypg.com

Replacing the fuse – EXFO FLS-2600B Tunable Laser Source User Manual

Page 58

background image

Maintenance

50

FLS-2600B

Replacing the Fuse

Replacing the Fuse

The FLS-2600B contains two fuses of type IEC, 250 V, 2 A, fast blow
0.197 in x 0.787 in/5 mm x 20 mm. The fuse holder is located at the back of
the FLS-2600B, just beside the power inlet.

To replace the fuses:
1.
Unplug the power cord from the FLS-2600B.

2. Pull the fuse holder out of the FLS-2600B.

3. Check and replace the fuses if necessary.

4. Make sure the fuses are placed firmly in the holder prior to its

reinsertion.

5. Firmly push the holder into place.

Use only with a 250V fuse

Employer uniquement avec

un fusible de 250V