beautypg.com

Output:ase, Output:mode – EXFO FLS-2600B Tunable Laser Source User Manual

Page 100

background image

Remote Control

92

FLS-2600B

Specific Commands

OUTPut:ASE?

Description This query returns a value indicating whether the ASE option is

available on the tunable laser source.

Syntax

OUTP:ASE?

Response

A boolean value indicating whether the ASE option is available

on the FLS-2600B Tunable Laser Source:

0: the ASE option is not available

1: the ASE option is available

Note

The response will always be “not available”.

See also

OUTP:MODE, OUTP:MODE?, OUTP:STAT and OUTP:STAT?

OUTPut:MODE

Description This command is used to set the source mode (tunable or ASE).

Syntax

OUTP:MODE

Parameters The parameter is a boolean parameter representing

the source mode:

0: Tunable

1: ASE

Note

This command cannot be performed if the FLS-2600B is

performing a sweep. It will return an error if ASE is selected

because it is not available.

Example

OUTP:MODE 0

See also

OUTP:ASE?, OUTP:MODE?, OUTP:STAT and OUTP:STAT?