Setting Sweep Parameters

36

FLS-2600B

Setting the Cycle Options

Setting the Cycle Options

The cycle options allow you to specify settings in both continuous and
stepped sweep modes.

You can set a specific number of cycles for the sweep to perform.

To set the number of cycles:
1.
Press Sweep.

2. Turn the selection dial clockwise until CYCLES is displayed.

3. Press ENTER. The Edit marker starts blinking in the bottom left-hand

corner of the display.

4. Turn the selection dial clockwise or counterclockwise to select the

desired number of sweep cycles. You can select a number from
1 to 99. You can also select Continuous cycling by turning the selection
dial counterclockwise until you pass 1. The display will then indicate
Loop.

5. Press ENTER to validate your choice, and Sweep to exit the Sweep

menu.