beautypg.com

EXFO FLS-2600B Tunable Laser Source User Manual

Page 92

background image

Remote Control

84

FLS-2600B

General Commands

*ESR?

Description This query reads the contents of the Standard Event Status

Register (ESR) and clears it.

Syntax

*ESR?

Response

An integer between 0 and 255.

*IDN?

Description This query reads the system identification string.

Syntax

*IDN?

Response

“EXFO E.-O. Eng. FLS-2600B vx.xx”xxxxxxxx xxxxxxx, where

xxxxxxxx xxxxxxx is the serial number and vx.xx is the current

product version.

*LOK

Description This command is used to set the Remote Lockout programming

state.

Syntax

*LOK

Parameters The parameter is a boolean value indicating if the

tunable laser source is
“0” or “unlocked”
“1” or “locked”

Note

This command can only be used when working with RS-232

communication.