Remote Control

Tunable Laser Source

87

General Commands

*RST

Description The *RST command sets the instrument to reset settings

(factory settings stored in ROM). Pending *OPC and *OPC?

actions are canceled. The instrument becomes idle. The *RST

command clears the Input and Error queues. The following

settings remain unchanged:

Instrument interface address

Output queue

Service Request Enable register (SRE)

Standard Event Status Enable register (ESE)

Syntax

*RST

Note

For a complete list of default parameters, see Saving and

Recalling Configurations on page 25.
This command uses electronic parts that have a limited life

span (non-volatile memory). You must use this command with

moderation.