Maintenance

54

FLS-2600B

Adjusting Your Unit According to Wavelength

Adjusting Your Unit According to Wavelength

It is possible to introduce an offset in your Tunable Laser Source to correct
a wavelength deviation as measured by a reference wavelength meter.

Note: A calibrated wavelength meter is required to perform a user calibration on

your unit.

To adjust your module at a certain wavelength:
1.
Turn your Tunable Laser Source on.

2. Press Setup.

3. Turn the selection dial until ADJUST  is displayed. The default

pre-selected wavelength appears.

4. Press ENTER. The Edit marker starts blinking in the lower part of the

display.

5. Connect your FLS-2600B to a calibrated wavelength meter.

6. Turn the selection dial until you reach the wavelength value read on

the wavelength meter, then press ENTER to validate it.

Note: The Fine-tune feature is not available in this submenu.

Edit

nm