beautypg.com

EXFO FLS-2600B Tunable Laser Source User Manual

Page 146

background image

Remote Control

138

FLS-2600B

FLS-2600B Error and Warning Messages

61

A problem occurred with the

thermo-electric cooler.

Contact EXFO.

62

The Flash memory was not

detected by the tunable laser

source.

Contact EXFO.

63

The Flash reading was not

performed in the set delays.

Contact EXFO.

64

The EEPROM was not

detected.

Contact EXFO.

65

A checksum error was

detected by the Tunable

Laser Source.

Contact EXFO.

66

A problem occurred while

the motor was moving.

Contact EXFO.

67

The Tunable Laser Source

index was lost.

Restart the Tunable Laser

Source.

68

The Tunable Laser Source is

outside its temperature

settings.

Ensure that the ambient

temperature is within the

Tunable Laser Source’s range.

69

An FPGA problem was

detected.

Contact EXFO.

71

An error occurred during a

sweep.

Retry your sweep.

72

An EEPROM error has been

detected.

Contact EXFO.

100

A command could not be

added to the command pipe.

Avoid turning the selection dial

while the Tunable Laser Source

is performing a sweep.

Error or

warning

Number

Description

Recommended Action