beautypg.com

Fls-2600b error and warning messages – EXFO FLS-2600B Tunable Laser Source User Manual

Page 145

background image

Remote Control

Tunable Laser Source

137

FLS-2600B Error and Warning Messages

FLS-2600B Error and Warning Messages

Error or

warning

Number

Description

Recommended Action

11

Nulling was not performed

correctly.

Restart your Tunable Laser

Source.

12

The Tunable Laser Source

returns the wrong

identification.

Contact EXFO.

20

There was a communication

error with the module.

Restart your Tunable Laser

Source.

25

There was a checksum error

while reading the source’s

FIFO.

Restart your Tunable Laser

Source.

30

The command was not

accepted and has been lost.

Verify if program resumed after

having lost a command. Rerun

the program.

31

A setting sent to the tunable

laser source is wrong.

Review command before

sending it again.

32

Command sent while

tunable laser source is

active.

Wait until Tunable Laser Source

is done before sending a

command.

34

Too many commands sent at

the same time.

Wait until the Tunable Laser

Source is done before sending

more commands.

40

The FIFO in the Tunable

Laser Source is not ready for

reading.

Wait for the FIFO to be ready.

60

A problem occurred with the

current going to the laser.

Contact EXFO.