beautypg.com

System:version, Unit:wavelength – EXFO FLS-2600B Tunable Laser Source User Manual

Page 124

background image

Remote Control

116

FLS-2600B

Specific Commands

SYSTem:VERSion?

Description This query reads the FLS-2600B identification string.

Syntax

SYST:VERS?

Response

“EXFO E.-O. Eng. FLS-2600B vx.xx”xxxxxxxx xxxxxxx, where

xxxxxxxx xxxxxxx is the serial number and vx.xx is the current

product version.

Note

This query returns the same response as the *IDN? query.

UNIT:WAVelength

Description This command changes the spectral measurement unit (nm or

THz). This command cannot be used in Sweep mode.

Syntax

UNIT:WAV

Parameters The parameter is a boolean parameter representing the

new spectral measurement unit:

1 – THz

0 – nm

Note

This command cannot be performed if the FLS-2600B is

performing a sweep.

Example

UNIT:WAV 1

See also

SOUR:WAV, SOUR:WAV?, SOUR:WAV:LIM:HIGH?,

SOUR:WAV:LIM:LOW?, SOUR:WAV:LIM:STEP? and UNIT:WAV?