Remote Control

86

FLS-2600B

General Commands

*OPC?

Description This query puts an ASCII “1” in the output queue when the

content of the input queue has been processed. This query is

useful to prevent another command from processing until the

current command is complete.

Syntax

*OPC?

Response

“1”

*REM

Description This command is used to set the Remote programming state.

Syntax

*REM

Parameters The parameter can be “1” to set the Remote

programming state to Remote or “0” to set the Remote

progamming state to Local.

Note