Status:operation:bit<8|9>: condition – EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual

Page 600

background image

Referencia de instrucciones SCPI

594

FTB-5240S/S-P/BP

Product-Specific Commands—Description

The response represents the

current operation condition of the instrument.

The meaning of the response depends on the

value returned for bit .

bit <8>: When the returned value is 1, the

instrument is performing a measurement (trigger

system INITiated).

bit <9>: When the returned value is 1, the

instrument is performing an offset nulling and/or

a wavelength referencing

(CALibration:ZERO:AUTO?).

Example(s)

STAT? Must return READY

CAL:ZERO:AUTO ONCE

STAT:OPER:BIT9:COND? Keep resending this

query as long as the operation is not complete

(returned value is not 0).

See Also

:CALibration:ZERO:AUTO?

:INITiate[:IMMediate]

:INITiate:CONTinuous?

:STATus?

:STATus:OPERation:BIT<8|9>:

CONDition?