beautypg.com

EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual

Page 452

background image

Referencia de instrucciones SCPI

446

FTB-5240S/S-P/BP

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the parameter.

The parameter corresponds to a

valid distance in hertz from peak to center of

noise region.

The

CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce:FREQuency? MIN and

CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce:FREQuency? MAX queries can be used to

determine valid distance values.

Example(s)

CALC:WDM:CHAN:DEF "C_23",195.0 THZ

CALC:WDM:CHAN:SEL "C_23"

CALC:WDM:CHAN:NOIS:TYPE POLY5

CALC:WDM:CHAN:NOIS:DIST:FREQ 125.0 GHZ

CALC:WDM:CHAN:NOIS:DIST:FREQ? Returns

1.250000E+011

Notes

Custom noise measurement distance is applied

only if selected noise type is POLYnomial5.

See Also

:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce[:WAVelength]

:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce:FREQuency?

:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT

h:FREQuency

:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE

:CALCulate[1..n][:WDM]:CHANnel:SELect

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:

DISTance:FREQuency

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:DISTance:FREQuency

This manual is related to the following products: