Initiate:continuous – EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual
Page 551

Referencia de instrucciones SCPI
OSA
545
Product-Specific Commands—Description
:INITiate:CONTinuous?
Description
This query indicates if the trigger system is
continuously initiated or not.
At *RST, this value is set to off (disabled).
Syntax
:INITiate:CONTinuous?
Parameter(s)
None
Response Syntax
Response(s)
Continuous:
The response data syntax for
defined as a
element.
The
state of the trigger system continuous cycle.
0: continuous cycle is disabled.
1: continuous cycle is enabled.
Example(s)
INIT:CONT ON
INIT:CONT? Returns 1 (trigger system
continuously initiated)
Notes
An acquisition may still be in progress even if
INIT:CONT? returns 0. The command
STAT:OPER:BIT8:COND? shall be used to test
acquisition completion.
See Also
:INITiate[:IMMediate]
:INITiate:CONTinuous
:STATus?
:STATus:OPERation:BIT<8|9>:CONDition?