Product-specific commands—description, Abort – EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual
Page 375

Referencia de instrucciones SCPI
OSA
369
Product-Specific Commands—Description
Product-Specific Commands—Description
:ABORt
Description
This command resets the trigger system and
places all trigger sequences in the IDLE state. Any
trace acquisition that is in progress is aborted as
quickly as possible. The command is not
completed until the trigger sequence is in the
IDLE state.
This command is an event and has no associated
*RST condition or query form.
Syntax
:ABORt
Parameter(s)
None
Example(s)
ABOR
Notes
A call to ABORt only returns once acquisition is
completely stopped and instrument is ready for
new commands. For this reason, execution of
this command may take a few seconds.
For a continuously initiated acquisition
(INIT:CONT ON), calling ABORt will
automatically set it to OFF.
See Also
:INITiate[:IMMediate]
:INITiate:CONTinuous
:STATus
:STATus:OPERation:BIT<8|9>:CONDition?