EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual
Page 529
Referencia de instrucciones SCPI
OSA
523
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:OSNR:
BANDwidth|BWIDth[:RESolution]
Description
This command sets custom resolution
bandwidth value for WDM analysis OSNR
calculation.
At *RST, this value is set to 0.100 nm.
Syntax
:CALCulate[1..n][:WDM]:OSNR:BANDwidth|BWI
Dth[:RESolution]
>|MAXimum|MINimum|DEFault
Parameter(s)
Resolution:
The program data syntax for
defined as a
followed by an optional DATA> element. The allowed PROGRAM DATA> element is M. The MAXimum and DEFault are accepted on input. MINimum allows to set the instrument to the smallest supported value. MAXimum allows to set the instrument to the greatest supported value.