Calculate[1..n][:wdm]:channel: noise:auto – EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual

Page 448

background image

Referencia de instrucciones SCPI

442

FTB-5240S/S-P/BP

Product-Specific Commands—Description

Example(s)

CALC:WDM:CHAN:DEF "C_001",192.1750 THZ

CALC:WDM:CHAN:SEL "C_001"

CALC:WDM:CHAN:NOIS:AUTO ON

CALC:WDM:CHAN:NOIS:AUTO? Returns 1 (auto

noise enabled)

Notes

Auto noise is available only if software option

"InB" is active.

Auto noise is computed only if analysed trace

was acquired using PMMH averaging type.

See Also

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:

AUTO

:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE

:CALCulate[1..n][:WDM]:CHANnel:NOISe:AUTO?

:CALCulate[1..n][:WDM]:CHANnel:SELect

:SENSe[1..n]:AVERage:TYPE

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:AUTO