Referencia de instrucciones SCPI

OSA

397

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:AUTO:

NOISe:WIDTh[WAVelength]?

Description

This query returns the wavelength width of the

noise measurement region of the WDM analysis

default channel.

At *RST, the width of the default channel noise

measurement region is set to 100.0 GHz.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:

WIDTh[WAVelength]?[MAXimum|MINi

mum|DEFault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

MAXimum|MINimum|DEFault.

MINimum is used to retrieve the instrument's

smallest supported value.

MAXimum is used to retrieve the instrument's

greatest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax

This manual is related to the following products: