Referencia de instrucciones SCPI

OSA

379

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:AUTO:

NOISe:AUTO

Description

This command controls activation of i-InBand

noise measurement for WDM analysis default

channel.

At *RST, auto noise is set to off (disabled).

Syntax

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:

AUTO

Parameter(s)

Auto:
The program data syntax for is defined

as a element. The

special forms ON and OFF are accepted

on input for increased readability. ON

corresponds to 1 and OFF corresponds to 0.

The parameter corresponds to the new

state of auto noise measurement.

0 or OFF: disable default channel auto noise.

1 or ON: enable default channel auto noise.