beautypg.com

Calculate[1..n][:wdm]:channel: noise:auto – EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual

Page 447

background image

Referencia de instrucciones SCPI

OSA

441

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:AUTO

Description

This command controls activation of i-InBand

noise measurement for WDM analysis of the

selected channel.

At *RST, this value is not available.

At CALCulate[1..n][:WDM]:CHANnel[:DEFine],

this value is set to off (disabled).

Syntax

:CALCulate[1..n][:WDM]:CHANnel:NOISe:AUTO

Parameter(s)

Auto:
The program data syntax for is defined

as a element. The

special forms ON and OFF are accepted

on input for increased readability. ON

corresponds to 1 and OFF corresponds to 0.

The parameter corresponds to the new

state of auto noise measurement.

0 or OFF: disable selected channel auto noise.

1 or ON: enable selected channel auto noise.

This manual is related to the following products: