beautypg.com

EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual

Page 459

background image

Referencia de instrucciones SCPI

OSA

453

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:WIDTh:FREQuency

Description

This command sets the frequency width of the

noise measurement region of the selected WDM

analysis channel.

At *RST, this value is not available.

At CALCulate[1..n][:WDM]:CHANnel[:DEFine],

width of the noise measurement region is set to

100.0 GHz.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT

h:FREQuencyHZ]>|MA

Ximum|MINimum|DEFault

Parameter(s)

Width:
The program data syntax for is defined

as a element followed by an

optional element.

The allowed

element is HZ. The special forms

MINimum, MAXimum and DEFault are accepted

on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

This manual is related to the following products: