Referencia de instrucciones SCPI

OSA

519

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:DATA:OSNR:

MEAN?

Description

This query returns computed WDM analysis

global result for mean signal-to-noise ratio.

At *RST, this value is not available.

Syntax

:CALCulate[1..n][:WDM]:DATA:OSNR:MEAN?

Parameter(s)

None

Response Syntax

Response(s)

Mean:
The response data syntax for is defined

as a

element.

The response corresponds to the

computed mean signal-to-noise ratio.

Example(s)

UNIT:RAT DB

CALC:WDM:DATA:OSNR:MEAN? Returns

4.471000E+001

Notes

Special NAN (not a number) value

-2251799813685248 is returned if analysis result

could not be computed.

See Also

:CALCulate[1..n][:WDM]:DATA:OSNR:FLATness?

:CALCulate[1..n][:WDM]:DATA:SIGnalPower:FLA

Tness?

:CALCulate[1..n][:WDM]:DATA:SIGnalPower:ME

AN?

:CALCulate[1..n][:WDM]:DATA:TPOWer?