Referencia de instrucciones SCPI

OSA

427

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:

CENTer:FREQuency?

Description

This query returns the nominal center frequency

of the selected WDM analysis channel.

At *RST, this value is not available.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:CENTer:FREQ

uency?[MAXimum|MINimum|DEFault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

MAXimum|MINimum|DEFault.

MINimum is used to retrieve the instrument's

smallest supported value.

MAXimum is used to retrieve the instrument's

greatest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax