Calculate[1..n][:wdm]:data: signalpower:flatness – EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual
Page 526

Referencia de instrucciones SCPI
520
FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:DATA:
SIGnalPower:FLATness?
Description
This query returns computed WDM analysis
global result for signal power flatness.
At *RST, this value is not available.
Syntax
:CALCulate[1..n][:WDM]:DATA:SIGnalPower:FLA
Tness?
Parameter(s)
None
Response Syntax
Response(s)
Flatness:
The response data syntax for
defined as a
element.
The
computed signal power flatness.
Example(s)
UNIT:RAT DB
CALC:DATA:SIGP:FLAT? Returns 3.118000E+001
Notes
Special NAN (not a number) value
-2251799813685248 is returned if analysis result
could not be computed.
See Also
:CALCulate[1..n][:WDM]:DATA:OSNR:FLATness?
:CALCulate[1..n][:WDM]:DATA:OSNR:MEAN?
:CALCulate[1..n][:WDM]:DATA:SIGnalPower:ME
AN?
:CALCulate[1..n][:WDM]:DATA:TPOWer?