Section 3. Installation and Setup
ADV6600
Y S I
Environmental
Page 25
Parameter Default Operating
Range
Comments
NO
3
A
2.543
Not checked
Cl J
99.5
Not checked
Cl S
-0.195
-0.217 to -0.15
Cl A
2.543
Not checked
Turb Offset
0
-10 to 10
Turb A1
500
Turb M1
500
0.6 to 1.5
Range is ratio of M1 to A1
Turb A2
1000
Turb M2
1000
0.6 to 1.5
Range is ratio of (M2-M1) to
(A2-A1)
Chl Offset
0
-30 to 20
Chl A1
500
Chl M1
500
0.6 to 1.5
Range is ratio of M1 to A1
Chl A2
1000
Chl M2
1000
0.6 to 1.5
Range is ratio of (M2-M1) to
(A2-A1)
Rhod Offset
0
-10 to 10
Rhod A1
500
Rhod M1
500
0.6 to 1.5
Range is ratio of M1 to A1
Rhod A2
1000
Rhod M2
1000
0.6 to 1.5
Range is ratio of (M2-M1) to
(A2-A1)
From the Advanced menu, select 2-Setup to activate or deactivate Auto sleep RS-232. Activation
of this feature enables a power savings system within the sonde. When enabled, power is only
applied to the sensors during sampling or calibration. The most important aspect of this feature is
its effect on the dissolved oxygen protocol. For this reason, the feature should be activated for most
applications. The selection should only be deactivated if you are running spot sampling studies
where the Averaging Interval and Sample Interval are set to the same value (see later).
From the Advanced menu, select 3-Sensor to display and change user-configurable constants as
shown. Select the appropriate number to view these parameters.