Y s i – YSI ADV6600 User Manual
Page 168
Appendix B. Specifications
ADV6600
Y S I
Environmental
Page 158
Sensor Range Resolution
Accuracy
Rhodamine WT
0 to 200 µg/L as
true dye; 1,000
µg/L as dye tracer
0.1 µg/L
± 5 % of the reading or 1 ug/L,
whichever is greater
Nitrate-Nitrogen
0-200 mg/L-N
0.001 mg/L-N to
1 mg/L-N, range
dependent
± 10% of reading or 2 mg/L,
whichever is greater
Ammonium/
Ammonia-
Nitrogen
0-200 mg/L-N
0.001 mg/L-N to
1 mg/L-N, range
dependent
± 10% of reading or 2 mg/L,
whichever is greater
Chloride
0-1000 mg/L
0.001 mg/L to
1 mg/L, range
dependent
± 15% of reading or 5 mg/L,
whichever is greater
ADV Sampling Volume:
~ 0.25cc located ~ 6.8 cm from probe tip
Operating Medium:
Fresh, sea or polluted water
Operating Temperature:
-5 to +45
°C
Operating Depth:
0 to 200 feet (60 meters), sonde and all sensors except Nitrate,
Ammonia, and Chloride which are 0 to 50 feet (15 meters)
Storage Temperature:
-40 to +60
°C for sonde and all sensors except pH and pH/ORP
-20 to +60
°C for pH and pH/ORP sensors
Size:
4 in. diameter; 24.5 in. length ( 10.1cm x 62.5 cm)
Weight with Batteries and probe
guard installed:
8.8 lb. (4.0 kg)
Power:
8 Standard Alkaline C-cells, 72Wh capacity,
or External 12-15 VDC
Power Consumption:
0.92 W (continuous operation); 0.001 W (standby)
Battery Life
At least 30 days for the following conditions: 1 min averaging
interval; 15 min sample interval; 20 C; all water quality
sensors active
Computer Interface:
RS-232, SDI-12
Internal Memory:
4 Mb
Software:
IBM Compatible PC with Pentium processor or above
Windows
®
98, NT, 2000, ME, or XP
32 MB memory
5 MB available disk space
CD-ROM Drive
RS-232 Serial Port
Monitor capable of 800 x 600 resolution or better