Guidance and manufacturer’s declaration – electromagnetic immunity

The ADview device is intended for use in the electromagnetic environment specified below. The customer or
the user of the ADview device should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance

level

Electromagnetic environment - guidance








Conducted RF
IEC 61000-4-6

Radiated RF IEC
61000-4-3








3 Vrms
150 kHz to 80 MHz

3V/m
80 MHz to 2.5 GHz








3V

3 V/m

Portable and mobile RF communications
equipment should be used no closer to any part
of the ADview, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = [3.5/V

1

] √P


d = [3.5/E

1

] √P 80MHz to 800MHz

d = [7/E

1

] √P 800MHz to 2.5GHz

where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey

a

,

should be less than the compliance level in
each frequency range

b

.

Interference may occur in the vicinity of
equipment marked with the following symbol:

NOTE 1—At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2—These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and
people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM

radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured
field strength in the location in which the ADview device is used exceeds the applicable RF compliance level above, the
ADview device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the ADview device.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.