American Diagnostic Corporation (ADC) ADview Modular Diagnostic Station User Manual
Page 50
40
Status Code: 20
Problem: Hardware error
Solutions:
1. Take another measurement.
2. If this is unsuccessful, power down, then power up the unit using the power button on the
right side of the unit. If the error does not recur immediately, take another measurement.
3. If the error recurs, remove power from the unit (unplug any power supplies and remove
the battery), then reconnect the power. If the error does not recur immediately, take
another measurement.
4. If the error recurs, contact ADC or an authorized service center.
Troubleshooting - Pulse Oximetry Module
Status Code: 01
Problem: Measurement time-out. The measurement time exceeded the 10-minute
time limit.
Solution: Remove the sensor from the patient. Redisplay the last measurement prior to the
timeout, or take a new measurement by placing the sensor on the patient.
Status Code: 02
Problem: Poor sensor position (signal is inadequate for a reliable measurement)
Solution: Adjust position of sensor on patient by placing sensor on opposite hand or ear or
alternate site. Avoid fingers with nail polish or artificial nails.
Status Code: 05
Problem: The sensor has been disconnected from the device.
Solution: Reconnect the sensor. If you wish, you may leave the sensor disconnected as this
code is only displayed once at the time the sensor is disconnected.
Status Code: 10
Problem: Defective sensor
Solution: Replace the sensor and take a new measurement.
Status Code: 20
Problem: Hardware error
Solutions:
1. Take another measurement.
2. If this is unsuccessful, power down, then power up the unit using the power button on the
right side of the unit. If the error does not recur immediately, take another measurement.
3. If the error recurs, remove power from the unit (unplug any power supplies and remove
the battery), then reconnect the power. If the error does not recur immediately, take
another measurement.
4. If the error recurs, contact ADC or an authorized service center.