The ADview is not intended for continuous monitoring. Although the blood pressure
cuff and cable are defibrillator proof, the temperature probe and SpO

2

sensor are not. Do

not leave the device unattended while taking measurements on a patient.

Only use such accessories as are recommended for use with this device. A list of
recommended accessories is on pages 61-66.

Do not operate the ADview near flammable anesthetics or volatile vapors. An explosion
may result.

Compressing the pneumatic tubing may cause system errors.

Do not use the device if it has failed its diagnostic self test or if it displays a greater than
zero pressure with no cuff attached or a value of functional oxygen saturation or
temperature with no sensor attached.

Prevent water or other fluids from entering any connectors or vents on the device.
Should this happen, all connectors should be dried with warm air. Then check the
calibration of the device and operating functions before reusing.

Do not make repairs yourself. Equipment must be returned to ADC or authorized service
personnel for repairs. Substitution of a component different from that supplied may
result in measurement error.