APPENDICES

Specifications

Patient population: Adult and pediatric patients (age 3 and above).

Method of measurement: Oscillometric

Initial inflation pressure: 160mmHg +/- 20mmHg

Blood pressure range (mmHg): 60< Systolic BP< 270, 30< Diastolic BP< 170

Blood pressure accuracy: Measurements determined with this device are equivalent to those obtained by a

trained observer using the cuff/stethoscope auscultation method, within the limits prescribed by
the American National Standard, Manual, electronic or automated sphygmomanometers.

Blood pressure determination time: 30-40 seconds typical for Adult cuff

Pulse rate range: 30-200 bpm +/- 2% or +/- 3 bpm, whichever is greater

Temperature range: 86

°F (30.0°C) – 109.4°F (43.0°C)

Temperature accuracy: +/- 0.2

°F (+/-0.1°C)

Functional oxygen saturation range: 40-100%

Functional oxygen saturation accuracy: 70-100% +/- 2 digits

(Note: because pulse oximeter equipment measurements are statistically distributed, only about
two-thirds of pulse oximeter equipment measurements can be expected to fall within ± 2 digits of
the value measured by a CO-oximeter.)


Operating conditions: 10

°C (50°F) to 40°C (104°F) Less than 90% RH


Storage conditions: -20

°C (-4°F) to 55°C (131°F) Less than 90% RH


Power: External power supply, AC adapter (ADC item number 9000AC9V)

Calibration: Check once per year for BP and Temperature