MANAGING READINGS

Recalling the Last Set of Readings

To redisplay the last set of readings, depress the automatic button

on the BP module for more than two

seconds until the last reading is displayed. If your ADview has temperature and/or pulse oximetry modules,
the last set of readings includes these readings as well. If the last attempted reading resulted in an error
and/or warning, then this will be displayed. The device will display dashes if no reading is in memory, a
reading was aborted or, the previous BP was a Sphygmode measurement.

For the pulse oximetry module: In the event of the 10-minute measurement timeout, the module will
terminate the measurement and status code “01” will display on the main module. The last valid reading
recorded at the end of the ten-minute period will be the recalled reading.

Clearing the Last Set of Readings

To clear the values from the last automatic BP measurement and the accessories, press and hold the

automatic button

more than 5 seconds. Previous values will be displayed momentarily. Then the

display blanks. On redisplay you will see dashes for all the values that have been cleared.

Your ADview displays the most recent set of readings for two minutes. If patient
privacy is a concern, you can clear these readings from the display before collecting
vital signs from another patient.

Bluetooth Wireless Transmission of Readings

If your ADview has the optional Bluetooth wireless communication capability you may be able to transmit
the readings into your EMR/EHR system or communication network. Contact your IT administrator or
EMR/EHR vendor for support on configuring your ADview for Bluetooth wireless communication.