Compliance – American Diagnostic Corporation (ADC) ADview Modular Diagnostic Station User Manual

Page 59

background image

49

Safety systems: Independent hardware over-pressure circuit and redundant software overpressure

algorithm to limit cuff pressure to less than 330 mmHg. Independent hardware timing circuit and
redundant software timer algorithm to limit the duration of a blood pressure cycle to less than 180
seconds.

Dimensions: Length = 5.5 inches, Height = 11.5 inches, Width = 3.8 inches; Length =14.0 cm, Height =

29.2 cm, Width = 9.7 cm

Standards: UL60601-1, CAN/CSA C22.2 601-1

IEC 60601-1, IEC 60601-1-2 (EMC), IEC 60601-1-4, ISO 9919, AAMI SP10:2002, ASTM E
1112, EN 12470-3

Meets EN-1060-1, Specification for non-invasive sphygmomanometers – Part 1: General
requirements and EN 1060-3, Non-invasive Sphygmomanometers – Part 3. Supplementary
Requirements for Electro-Mechanical BP Measuring Systems”

Classification: Protection against electric shock: Class II (for non-battery version), Internally Powered

Equipment (for battery version); Applied parts: Type BF; Mode of operation: Continuous

IP index: IPX0

Compliance

American Diagnostic Corporation
Oakfield Industrial Estate
Eynsham, Oxfordshire OX29 4TS
UK
Tel: +44. 1865.884.234
Fax: +44. 1865.884.235

Safety Requirements

Clinical grade BP measurement accuracy defined by fully meeting the requirements of:

AAMI SP-10 2002

EN 1060-4

EMC Statement